PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-47030
- Event Type
- Malfunction
- Date Received
- November 4, 2014
- Date of Event
- October 6, 2014
- Report Date
- October 6, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
ALL OF THE BUTTONS FUNCTIONED PROPERLY. HOWEVER, MOISTURE DAMAGE WAS NOTED ON THE KEYPAD TRACES. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT THE DISPLAY WINDOW CORNERS, A CRACKED RESERVOIR TUBE LIP AND CRACKED BATTERY TUBE THREADS.
IT WAS REPORTED THAT THE CUSTOMER HAD HIGH BLOOD GLUCOSE LEVELS OF 183 MG/DL. IT WAS REPORTED THAT SOME OF THE BUTTONS OF THE INSULIN PUMP WERE UNRESPONSIVE. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WOULD NEED TO BE REPLACED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO A BACK-UP PLAN PER THE HEALTH CARE PROVIDER'S INSTRUCTION. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 707138 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR |