FDA Adverse Event Injury Summary report: N

ICHEM VELOCITY

MDR report key: 4223499 · Received October 29, 2014

Report

Report Number
2023446-2014-00173
Event Type
Injury
Date Received
October 29, 2014
Date of Event
October 6, 2014
Report Date
October 6, 2013
Manufacturer
IRIS INTERNATIONAL
Product Code
KQO
PMA / PMN Number
K101852
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A LARGE NOISE AND THEN A BURNING SMELL FROM THE VELOCITY WAS REPORTED. AN IRIS FIELD SERVICE ENGINEER (FSE) OBSERVED THE BURNT SMELL FROM INSIDE THE ANALYZER CAUSED BY THE BURNT COMPONENTS FROM THE POWER SUPPLY. FSE REPLACED THE POWER SUPPLY AND THE ISSUE WAS RESOLVED. THERE WERE NO REPORTS OF INJURIES, VISIBLE FLAMES, SMOKE EMISSION, AND THE FIRE DEPARTMENT WAS NOT CALLED.

Description of Event or Problem · 1

CUSTOMER REPORTED BURNING SMELL FROM THE UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
692770 ICHEM VELOCITY AUTOMATED URINE CHEMISTRY ANALYZER KQO IRIS INTERNATIONAL

Patients

Seq Age Sex Outcome Treatment
1 Other