FDA Adverse Event
Injury
Summary report: N
ICHEM VELOCITY
MDR report key: 4223499
·
Received October 29, 2014
Report
- Report Number
- 2023446-2014-00173
- Event Type
- Injury
- Date Received
- October 29, 2014
- Date of Event
- October 6, 2014
- Report Date
- October 6, 2013
- Manufacturer
- IRIS INTERNATIONAL
- Product Code
- KQO
- PMA / PMN Number
- K101852
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A LARGE NOISE AND THEN A BURNING SMELL FROM THE VELOCITY WAS REPORTED. AN IRIS FIELD SERVICE ENGINEER (FSE) OBSERVED THE BURNT SMELL FROM INSIDE THE ANALYZER CAUSED BY THE BURNT COMPONENTS FROM THE POWER SUPPLY. FSE REPLACED THE POWER SUPPLY AND THE ISSUE WAS RESOLVED. THERE WERE NO REPORTS OF INJURIES, VISIBLE FLAMES, SMOKE EMISSION, AND THE FIRE DEPARTMENT WAS NOT CALLED.
Description of Event or Problem · 1
CUSTOMER REPORTED BURNING SMELL FROM THE UNIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 692770 | ICHEM VELOCITY | AUTOMATED URINE CHEMISTRY ANALYZER | KQO | IRIS INTERNATIONAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |