MDR report key: 4223437 · Received August 6, 2014

Report

Report Number
1720753-2014-06758
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
June 30, 2014
Report Date
August 6, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. A COMPLETE FILAMENT CALIBRATION WAS PERFORMED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A FILAMENT REGULATOR ERROR. THE FIELD ENGINEER RESPONDED THAT THIS ISSUE DID HAPPEN DURING A CASE. THIS MALFUNCTION MAY HAVE A RESULTED IN A POSSIBLE ARO (ACCIDENTAL RADIATION OCCURRENCE). THERE IS NO REPORT OF INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
459336 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment