FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 4223437
·
Received August 6, 2014
Report
- Report Number
- 1720753-2014-06758
- Event Type
- Malfunction
- Date Received
- August 6, 2014
- Date of Event
- June 30, 2014
- Report Date
- August 6, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. A COMPLETE FILAMENT CALIBRATION WAS PERFORMED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A FILAMENT REGULATOR ERROR. THE FIELD ENGINEER RESPONDED THAT THIS ISSUE DID HAPPEN DURING A CASE. THIS MALFUNCTION MAY HAVE A RESULTED IN A POSSIBLE ARO (ACCIDENTAL RADIATION OCCURRENCE). THERE IS NO REPORT OF INJURY OR DEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 459336 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |