FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4223430 · Received November 4, 2014

Report

Report Number
2032227-2014-46991
Event Type
Malfunction
Date Received
November 4, 2014
Date of Event
July 5, 2014
Report Date
October 5, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP ALARMED FAILED BATTERY TEST, DUE TO CORRODED BATTERY TUBE AND BATTERY CAP CONTACT. INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTONS, DUE TO CORRODED KEYPAD TRACES. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON LCD WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP HAD A NONRESPONSIVE KEYPAD. THE CUSTOMER ALSO HAD MULTIPLE FAILED BATTERY TESTS. THE CUSTOMER'S BLOOD GLUCOSE LEVELS ARE UNKNOWN. STANDARD TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. THE PRODUCT HAS BEEN RETURNED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
707578 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAL

Patients

Seq Age Sex Outcome Treatment
1 22 YR