FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4223362 · Received November 4, 2014

Report

Report Number
2032227-2014-46830
Event Type
Malfunction
Date Received
November 4, 2014
Date of Event
October 7, 2014
Report Date
October 7, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FINDINGS: UNABLE TO PERFORM OPERATING CURRENTS, SELF TEST, BASIC OCCLUSION, OCCLUSION AND EXCESSIVE NO DELIVERY TEST DUE TO BLANK DISPLAY. UNIT RECEIVED WITH BLANK DISPLAY DUE TO BROKEN SOLDER JOINT B1 I/B. PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW. UNIT RECEIVED CRACKED CASE DISPLAY WINDOW CORNER. UNIT RECEIVED WITH CRACKED BATTERY TUBE THREADS. PUMP RECEIVED WITH BROKEN RESERVOIR TUBE LIP. (B)(4). THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP NEEDS A COMPLETE FUNCTIONAL ANALYSIS. THE BLOOD GLUCOSE LEVEL WAS NOT INCLUDED IN THE REPORT. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
707447 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-554WWS

Patients

Seq Age Sex Outcome Treatment
1