FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4223359 · Received November 4, 2014

Report

Report Number
2032227-2014-46831
Event Type
Malfunction
Date Received
November 4, 2014
Date of Event
October 9, 2014
Report Date
October 9, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FINDINGS: PUMP WAS RECEIVED WITH MOTOR ERROR ALARM DURING OCCLUSION TEST AND UNABLE TO PRIME AT 2.5 LBS DURING PRIME/A33 TEST DUE TO FAULTY FSR (GOLD). MOTOR PASSED MOTOR TEST. UNIT HAD MINOR SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.(B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP FAILED THE SELF TEST. CUSTOMER'S CURRENT BLOOD GLUCOSE READING WAS 241 MG/DL. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
707446 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAL

Patients

Seq Age Sex Outcome Treatment
1 33 YR