FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4223359
·
Received November 4, 2014
Report
- Report Number
- 2032227-2014-46831
- Event Type
- Malfunction
- Date Received
- November 4, 2014
- Date of Event
- October 9, 2014
- Report Date
- October 9, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FINDINGS: PUMP WAS RECEIVED WITH MOTOR ERROR ALARM DURING OCCLUSION TEST AND UNABLE TO PRIME AT 2.5 LBS DURING PRIME/A33 TEST DUE TO FAULTY FSR (GOLD). MOTOR PASSED MOTOR TEST. UNIT HAD MINOR SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.(B)(4)
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP FAILED THE SELF TEST. CUSTOMER'S CURRENT BLOOD GLUCOSE READING WAS 241 MG/DL. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 707446 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR |