FDA Adverse Event Malfunction Summary report: N

PERMANENT CAUTERY SPATULA INSTRUMENT

MDR report key: 4223353 · Received November 4, 2014

Report

Report Number
2955842-2014-05585
Event Type
Malfunction
Date Received
November 4, 2014
Date of Event
August 25, 2014
Report Date
October 10, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND A BROKEN PITCH UP CABLE AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY WEAR MARKS. ISI HAS CONDUCTED A DEVICE HISTORY RECORD (DHR) REVIEW FOR THIS DEVICE AND DID NOT FIND ANY NON-CONFORMANCES THAT WERE RELATED TO THIS REPORTED EVENT. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE BROKEN PITCH CABLE FOUND DURING FAILURE ANALYSIS COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, A BROKEN CABLE WAS IDENTIFIED ON A PERMANENT CAUTERY SPATULA INSTRUMENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED. THERE WERE NO REPORTS OF FRAGMENTS FALLING INTO THE PATIENT.