PERMANENT CAUTERY SPATULA INSTRUMENT
Report
- Report Number
- 2955842-2014-05585
- Event Type
- Malfunction
- Date Received
- November 4, 2014
- Date of Event
- August 25, 2014
- Report Date
- October 10, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND A BROKEN PITCH UP CABLE AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY WEAR MARKS. ISI HAS CONDUCTED A DEVICE HISTORY RECORD (DHR) REVIEW FOR THIS DEVICE AND DID NOT FIND ANY NON-CONFORMANCES THAT WERE RELATED TO THIS REPORTED EVENT. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE BROKEN PITCH CABLE FOUND DURING FAILURE ANALYSIS COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.
IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, A BROKEN CABLE WAS IDENTIFIED ON A PERMANENT CAUTERY SPATULA INSTRUMENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED. THERE WERE NO REPORTS OF FRAGMENTS FALLING INTO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 707444 | PERMANENT CAUTERY SPATULA INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420184-06 | M10140516 421 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |