FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4223236 · Received November 4, 2014

Report

Report Number
2032227-2014-47405
Event Type
Injury
Date Received
November 4, 2014
Date of Event
September 30, 2014
Report Date
October 6, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4). THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

BOLUSES WERE DELIVERED MANUALLY AND THE RESULTS SHOWED UP PROPERLY IN THE BOLUS HISTORY AND DAILY TOTALS SCREEN. A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP WAS RECEIVED WITH A CRACKED DISPLAY WINDOW, CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, SCRATCHED RESERVOIR TUBE WINDOW, CRACKED BATTERY TUBE THREADS, A CRACKED BELT CLIP SLOT AND A MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED A HIGH BLOOD GLUCOSE LEVEL OF 472 MG/DL AND A LOW BLOOD GLUCOSE LEVEL OF 38 MG/DL. THE CUSTOMER COLLAPSED FROM THE LOW BLOOD GLUCOSE LEVEL, BUT DID NOT GO TO THE HOSPITAL. THE CUSTOMER DECLINED TROUBLESHOOTING. THE PRODUCT HAS BEEN RETURNED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
706099 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-554CAP

Patients

Seq Age Sex Outcome Treatment
1 Other