FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4223194 · Received November 4, 2014

Report

Report Number
2531779-2014-31390
Event Type
Malfunction
Date Received
November 4, 2014
Report Date
October 20, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 12/31/2014 - DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/10/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX DATA REVEALED NO EVIDENCE OF SHORT BATTERY LIFE. THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED, AND THE BATTERY CAP HAD DAMAGED THREADS. THE BATTERY CAP WAS UNABLE TO SECURE ONTO THE PUMP. THE PUMP POWERED ON WITH THE RETURNED BATTERY CAP. THE ELECTRICAL CURRENT DRAWS MEASURED WITHIN SPECIFICATIONS. THE PUMP WAS OPENED AND NO DAMAGE WAS OBSERVED. UNRELATED TO THE COMPLAINT, THE DISPLAY SCREEN WAS DIM AND DISCOLORED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE W/ DAMAGE) ISSUE. THE REPORTER ALLEGED THAT THE BATTERY COMPARTMENT WAS CRACKED, AND THAT THE BATTERY CAP HAD NOT BEEN REPLACED IN FOUR TO SIX MONTHS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
705479 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 15 YR