PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-47188
- Event Type
- Malfunction
- Date Received
- November 4, 2014
- Date of Event
- October 6, 2014
- Report Date
- October 6, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
THE INSULIN PUMP PASSED THE BASIC OCCLUSION TEST AND THE DISPLACEMENT TEST. NO MOTOR ERROR ALARMS WERE NOTED. HOWEVER, THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST DUE TO A FAULTY FORCE SENSOR RESISTOR. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT THE DISPLAY WINDOW CORNERS AND CRACKED BATTERY TUBE THREADS.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).
CUSTOMER REPORTED A MOTOR ERROR ALARM DURING A PRIME. CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. HE ALSO STATED HE WAS ABLE TO COMPLETE THE REWIND SEQUENCE ON THE INSULIN PUMP. CUSTOMER ALSO STATED HE HAD CHANGED HIS INFUSION SET TWICE, BUT THE ALARM PERSISTS. IT WAS ALSO FOUND INSULIN WAS ABLE TO EXIT WITH A PLUNGER DURING A MANUAL PRIME. THE CUSTOMER COULD NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ALARM. THE CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 705627 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 9 YR |