ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2014-31388
- Event Type
- Malfunction
- Date Received
- November 4, 2014
- Report Date
- October 20, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
DEVICE EVALUATION: THE RETAINED CARTRIDGE WAS EVALUATED BY PRODUCT ANALYSIS ON 11/17/2014 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE RETAINED CARTRIDGE PASSED VISUAL INSPECTION WITH NO DAMAGE OR DEFECTS NOTED. A FILL TEST WAS COMPLETED WITH NO AIR BUBBLES BEING FORMED INSIDE THE CARTRIDGE. A LEAK TEST WAS PERFORMED WITH NO FAILURES BEING OBSERVED; NO LEAKS WERE OBSERVED FROM THE LUER CONNECTION, O-RINGS, OR ANYWHERE ELSE IN THE CARTRIDGE. THERE WAS NO DEFECT FOUND.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 705630 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | C200010 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |