FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4223098 · Received November 4, 2014

Report

Report Number
2032227-2014-47414
Event Type
Malfunction
Date Received
November 4, 2014
Date of Event
December 19, 2013
Report Date
December 19, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

PUMP PASSED A21 ALARM TEST. NO A21 ALARMS OR A17 ALARMS NOTED. NO A47 ALARMS NOTED DURING TESTING. HOWEVER A47 ALARMS NOTED IN ALARM HISTORY DUE TO CORRUPTED HISTORY FILES. UNABLE TO DOWNLOAD PROPERLY DUE TO A47 ALARMS. CRACKED RESERVOIR TUBE LIP AND MINOR SCRATCHES ON DISPLAY WINDOW NOTED DURING VISUAL INSPECTION.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP WAS NOT COMPLETING THE UPLOAD PROCESS TO CARELINK. CUSTOMER REPORTED THAT THE INSULIN PUMP'S ALARM HISTORY SHOWED MULTIPLE ERROR ALARMS. BLOOD GLUCOSE LEVEL WAS NOT PROVIDED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
706225 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAL

Patients

Seq Age Sex Outcome Treatment
1 46 YR