FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4223095 · Received November 4, 2014

Report

Report Number
2032227-2014-47410
Event Type
Malfunction
Date Received
November 4, 2014
Date of Event
October 17, 2011
Report Date
October 17, 2011
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, SCRATCHED DISPLAY WINDOW, CRACKED RESERVOIR TUBE, CRACKED RESERVOIR TUBE WINDOW AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION. A21 AFTER BATTERY CHANGE DUE TO BROKEN SOLDER JOINT C13 ON I/B. NO UNEXPECTED BATTERY OUT LIMIT ALARM NOTED. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP RETURNED MULTIPLE ERROR ALARMS DURING BASAL. CUSTOMER ALSO REPORTED RECEIVING A BATTERY OUT LIMIT EACH TIME SHE CHANGED THE INSULIN PUMP'S BATTERY. BLOOD GLUCOSE LEVEL WAS NOT PROVIDED. NO FURTHER INFORMATION WAS PROVIDED.