FDA Adverse Event Malfunction Summary report: N

RESOLUTION CLIP CLIPPING DEVICE

MDR report key: 4222744 · Received November 4, 2014

Report

Report Number
3005099803-2014-03499
Event Type
Malfunction
Date Received
November 4, 2014
Date of Event
October 8, 2014
Report Date
October 14, 2014
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
KOG
PMA / PMN Number
K040148
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ALTHOUGH THE SUSPECT DEVICE HAS BEEN RECEIVED, THE EVALUATION HAS NOT BEEN COMPLETED. THEREFORE, THE CAUSE OF THE REPORTED MALFUNCTION HAS NOT BEEN DETERMINED. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MDR WILL BE FILED.

Additional Manufacturer Narrative · 1

A VISUAL EXAMINATION OF THE RETURNED DEVICE NOTED THAT THE DEVICE WAS HALF DEPLOYED. THE YOKE WHICH WAS STILL ATTACHED TO THE CONTROL WIRE WAS ACTUATED AND WAS ABLE TO BE DEPLOYED. THE CLIP ASSEMBLY WAS NOT RETURNED. GIVEN THE EVENT DESCRIPTION AND CONDITION OF THE RETURNED DEVICE, THERE ISN'T ENOUGH INFORMATION TO DETERMINE A PROBABLE ROOT CAUSE. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY, AND PRODUCT SPECIFICATIONS AT THE TIME OF RELEASE TO DISTRIBUTION.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2014 THAT A RESOLUTION CLIP DEVICE WAS USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CLIP GRASPED AND LOCKED ONTO TISSUE; HOWEVER, THE CLIP WOULD NOT RELEASE FROM THE CATHETER TO DEPLOY. AFTER MANIPULATION, THE CLIP SUCCESSFULLY RELEASED FROM THE CATHETER. ANOTHER RESOLUTION CLIP WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2014 THAT A RESOLUTION CLIP DEVICE WAS USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CLIP GRASPED AND LOCKED ONTO TISSUE; HOWEVER, THE CLIP WOULD NOT RELEASE FROM THE CATHETER TO DEPLOY. AFTER MANIPULATION, THE CLIP SUCCESSFULLY RELEASED FROM THE CATHETER. ANOTHER RESOLUTION CLIP WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.