RESOLUTION CLIP CLIPPING DEVICE
Report
- Report Number
- 3005099803-2014-03499
- Event Type
- Malfunction
- Date Received
- November 4, 2014
- Date of Event
- October 8, 2014
- Report Date
- October 14, 2014
- Manufacturer
- BOSTON SCIENTIFIC - MARLBOROUGH
- Product Code
- KOG
- PMA / PMN Number
- K040148
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). ALTHOUGH THE SUSPECT DEVICE HAS BEEN RECEIVED, THE EVALUATION HAS NOT BEEN COMPLETED. THEREFORE, THE CAUSE OF THE REPORTED MALFUNCTION HAS NOT BEEN DETERMINED. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MDR WILL BE FILED.
A VISUAL EXAMINATION OF THE RETURNED DEVICE NOTED THAT THE DEVICE WAS HALF DEPLOYED. THE YOKE WHICH WAS STILL ATTACHED TO THE CONTROL WIRE WAS ACTUATED AND WAS ABLE TO BE DEPLOYED. THE CLIP ASSEMBLY WAS NOT RETURNED. GIVEN THE EVENT DESCRIPTION AND CONDITION OF THE RETURNED DEVICE, THERE ISN'T ENOUGH INFORMATION TO DETERMINE A PROBABLE ROOT CAUSE. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY, AND PRODUCT SPECIFICATIONS AT THE TIME OF RELEASE TO DISTRIBUTION.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2014 THAT A RESOLUTION CLIP DEVICE WAS USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CLIP GRASPED AND LOCKED ONTO TISSUE; HOWEVER, THE CLIP WOULD NOT RELEASE FROM THE CATHETER TO DEPLOY. AFTER MANIPULATION, THE CLIP SUCCESSFULLY RELEASED FROM THE CATHETER. ANOTHER RESOLUTION CLIP WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2014 THAT A RESOLUTION CLIP DEVICE WAS USED DURING A COLONOSCOPY PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE CLIP GRASPED AND LOCKED ONTO TISSUE; HOWEVER, THE CLIP WOULD NOT RELEASE FROM THE CATHETER TO DEPLOY. AFTER MANIPULATION, THE CLIP SUCCESSFULLY RELEASED FROM THE CATHETER. ANOTHER RESOLUTION CLIP WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.