FDA Adverse Event Malfunction Summary report: N

SCREWDRIVER BIT AXSOS T15, AO FITTING 4.0MM LOCKING SET

MDR report key: 4222403 · Received November 4, 2014

Report

Report Number
0008031020-2014-00534
Event Type
Malfunction
Date Received
November 4, 2014
Date of Event
October 11, 2014
Report Date
October 11, 2014
Manufacturer
STRYKER TRAUMA SELZACH
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE REPORTED EVENT THAT THE TIP OF THE SCREWDRIVER BLADE BROKE OFF COULD BE CONFIRMED. THE DEVICE HAS BEEN RECEIVED INCOMPLETE AND OBVIOUSLY DAMAGED. THE TIP WAS BROKEN AND THE FRAGMENT HAS NOT BEEN RECEIVED FOR INVESTIGATION. THE FRACTURE OCCURRED DIAGONALLY ACROSS THE TORX SHAPED TIP OF THE SCREW DRIVER BLADE AND WAS CAUSED BY EXCESSIVE FORCE, WHICH MOST LIKELY WAS APPLIED BY A POWER TOOL. BASED ON THE INVESTIGATION, THE ROOT CAUSE OF THE REPORTED BROKEN BLADE WAS ATTRIBUTED TO A USER RELATED ISSUE. A HANDLING TEST FOR T15 SCREWDRIVER AXSOS BLADE BREAKAGE STATES THAT THE TORQUE TO FAILURE OF THE TESTED BLADES, CATALOG NUMBER 702753, WAS AT LEAST 7.297NM, AND THAT SCREWDRIVER BLADES COULD NOT BE BROKEN DURING MANUAL SCREW INSERTION, AS IT IS DESCRIBED IN THE OPERATIVE TECHNIQUE. THE UNIVERSAL TORQUE LIMITING ATTACHMENT, 4NM, AO FITTING, 4.0MM LOCKING SET, CATALOG NUMBER 702750, WHICH IS LISTED IN THE OPERATIVE TECHNIQUE FOR PROXIMAL HUMERUS (LITERATURE NUMBER : 982275), LIMITS THE TORQUE TO 4NM. THE OPERATIVE TECHNIQUE CONTAINS THE FOLLOWING STATEMENT: " ADDITIONAL TIPS: [¿] 5. DO NOT USE POWER FOR FINAL INSERTION OF LOCKING SCREWS. IT IS IMPERATIVE TO ENGAGE THE SCREW HEAD INTO THE PLATE USING THE TORQUE LIMITING ATTACHMENT. ENSURE THAT THE SCREWDRIVER TIP IS FULLY SEATED IN THE SCREW HEAD, BUT DO NOT APPLY AXIAL FORCE DURING FINAL TIGHTENING. IF THE SCREW STOPS SHORT OF FINAL POSITION, BACK UP A FEW TURNS AND ADVANCE THE SCREW AGAIN (WITH TORQUE LIMITER ON). " A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. NO INDICATIONS OF ANY MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE DETERMINED DURING THE INVESTIGATION.

Description of Event or Problem · 1

IT WAS REPORTED DURING PRIMARY SURGERY OF THE PROXIMAL HUMERUS, THE DOCTOR WAS SCREWING A 4.0 LOCKING SCREW, AND THOUGHT THE SCREWDRIVER WAS STRIPPING, BUT IT TURNED OUT THAT THE TIP OF THE SCREWDRIVER BROKE OFF. THE DOCTOR THEN USED ANOTHER ONE TO COMPLETE THE SURGERY. NO DELAY.

Description of Event or Problem · 1

IT WAS REPORTED DURING PRIMARY SURGERY OF THE PROXIMAL HUMERUS, THE DOCTOR WAS SCREWING A 4.0 LOCKING SCREW, AND THOUGHT THE SCREWDRIVER WAS STRIPPING, BUT IT TURNED OUT THAT THE TIP OF THE SCREWDRIVER BROKE OFF. THE DOCTOR THEN USED ANOTHER ONE TO COMPLETE THE SURGERY. NO DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
708258 SCREWDRIVER BIT AXSOS T15, AO FITTING 4.0MM LOCKING SET ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LXH STRYKER TRAUMA SELZACH 01208Y

Patients

Seq Age Sex Outcome Treatment
1 74 YR Other