FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 4222293
·
Received November 3, 2014
Report
- Report Number
- 2531779-2014-31359
- Event Type
- Malfunction
- Date Received
- November 3, 2014
- Report Date
- October 20, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/20/2014 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE PUMP WAS POWERED ON AND THE DISPLAY SCREEN APPEARED DIM AND DISCOLORED. UNRELATED TO THE DISPLAY ISSUE, THE KEYPAD COVER WAS PEELING AT THE OK AND DOWN ARROW BUTTONS. ALL OF THE KEYPAD BUTTONS RESPONDED PROPERLY TO USER PRESSES. (B)(4).
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DIM AND DISCOLORED DISPLAY. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 10/20/2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 702609 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 22 YR |