FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 4222262 · Received November 3, 2014

Report

Report Number
2531779-2014-31335
Event Type
Malfunction
Date Received
November 3, 2014
Report Date
October 20, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 01/08/2015 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DURING TESTING, LIGHT PRESSURE WAS APPLIED TO THE U39 COMPONENT ON THE PRINTED CIRCUIT BOARD, AND THE PUMP EMITTED A CS 069 CALL SERVICE ALARM. THE CALL SERVICE ALARM WAS RECORDED IN THE BLACK BOX DATA AS A CS 087 FAILURE, INDICATING A FAILURE OF THE COMPONENT. UNRELATED TO THE COMPLAINT, THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. IT WAS REPORTED THAT THE PUMP EMITTED A CS 069 CALL SERVICE ALARM. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
702956 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1