FDA Adverse Event
Malfunction
Summary report: N
PARADIGM INSULIN INFUSION PUMP
MDR report key: 4222137
·
Received November 3, 2014
Report
- Report Number
- 2032227-2014-46845
- Event Type
- Malfunction
- Date Received
- November 3, 2014
- Date of Event
- June 16, 2014
- Report Date
- June 16, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- LZG
- PMA / PMN Number
- K040676
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FINDINGS: UNIT RECEIVED MOTOR ERROR ALARM DURING REWIND DUE TO ENCODER OUT OF PHASE. UNABLE TO PERFORM DISPLACEMENT TEST DUE TO CONSTANT MOTOR ERROR ALARM. PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW. UNIT RECEIVED CRACKED CASE DISPLAY WINDOW CORNER. UNIT RECEIVED WITH CRACKED BATTERY TUBE THREADS. PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER HAD A MOTOR ERROR ALARM ON THE INSULIN PUMP. ALARM OCCURRED MORE THAN 3 TIMES. NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 702574 | PARADIGM INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | MEDTRONIC MINIMED | MMT-715WWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |