FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 4222137 · Received November 3, 2014

Report

Report Number
2032227-2014-46845
Event Type
Malfunction
Date Received
November 3, 2014
Date of Event
June 16, 2014
Report Date
June 16, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
LZG
PMA / PMN Number
K040676
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FINDINGS: UNIT RECEIVED MOTOR ERROR ALARM DURING REWIND DUE TO ENCODER OUT OF PHASE. UNABLE TO PERFORM DISPLACEMENT TEST DUE TO CONSTANT MOTOR ERROR ALARM. PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW. UNIT RECEIVED CRACKED CASE DISPLAY WINDOW CORNER. UNIT RECEIVED WITH CRACKED BATTERY TUBE THREADS. PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD A MOTOR ERROR ALARM ON THE INSULIN PUMP. ALARM OCCURRED MORE THAN 3 TIMES. NO ADDITIONAL INFORMATION PROVIDED.