RESERVOIR 3ML
Report
- Report Number
- 2032227-2014-46689
- Event Type
- Malfunction
- Date Received
- November 3, 2014
- Date of Event
- October 4, 2014
- Report Date
- October 4, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K032005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
INSPECTED RESERVOIR, SNAP-CAP AND SEPTUM FOR ANOMALIES NONE WERE FOUND. PERFORMED OCCLUSION TEST PER SPECIFICATIONS, RESERVOIR FILLED AND CONNECTED TO NEW INFUSION SET. INSTALLED RESERVOIR AND CONNECTOR INTO INSULIN PUMP AND PERFORMED INFUSION SET. INSTALLED RESERVOIR AND CONNECTOR TO INSULIN PUMP AND PERFORMED CATHETER TIP. RESERVOIR NOT OCCLUDED.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4) SITE PER VARIANCE 5.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED NO DELIVERY WHILE ATTEMPTING TO FILL THE TUBING. THE BLOOD GLUCOSE READING WAS UNKNOWN. THE CUSTOMER WANT TO RETURN THE INFUSION SET AND RESERVOIR FOR ANALYSIS. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 703657 | RESERVOIR 3ML | RESERVOIR | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-332A | HG04680 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR |