FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 4221752 · Received November 3, 2014

Report

Report Number
2032227-2014-46684
Event Type
Malfunction
Date Received
November 3, 2014
Date of Event
October 4, 2014
Report Date
October 4, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP PASSED DISPLACEMENT TEST, REWIND, BASIC OCCLUSION AND PRIME TESTS. INSULIN PUMP WAS RECEIVED WITH STUCK IN MOTOR ERROR ALARM LOOP DURING BOLUS AND BASAL DELIVERY. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE MOTOR MAY HAVE HAD AN INTERMITTENT FAILURE THAT WAS NOT DETECTED DURING OUR TESTING. INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING A BOLUS ATTEMPT. THE BLOOD GLUCOSE READING WAS 100 MG/DL. THE CUSTOMER WAS UNABLE TO COMPLETE THE REWIND SEQUENCE. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
702678 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAS

Patients

Seq Age Sex Outcome Treatment
1 49 YR