G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2014-27932
- Event Type
- Malfunction
- Date Received
- November 3, 2014
- Date of Event
- October 7, 2014
- Report Date
- October 7, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
(B)(4). THE RETURNED COMPLAINT DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS ATTEMPTED, HOWEVER, THE DEVICE WOULD NOT TURN ON. A REVIEW OF THE RECEIVER DATA LOG CONFIRMED A HARDWARE AND A FIRMWARE ERROR CODE. THE ROOT CAUSE WAS DETERMINED TO BE A HARDWARE COMPONENT FAILURE. CAPA HAS BEEN INITIATED FOR THIS ISSUE.
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 AND CLAIMED THAT ON (B)(6) 2014 THE PATIENT EXPERIENCED A HARDWARE FAILURE. DEXCOM TECHNICAL SUPPORT ADVISED THE PATIENT TO RESET THE DEVICE ON (B)(6) 2014. DEXCOM HAS ISSUED AN RGA FOR PATIENT TO RETURN THE DEVICE TO BE INVESTIGATED. PATIENT DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION AT THE TIME OF CONTACT WITH DEXCOM TECHNICAL SUPPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 703215 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | MT20649-PNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR |