FDA Adverse Event Malfunction Summary report: N

ABSORB BIORESORBABLE VASCULAR SCAFFOLD SYSTEM

MDR report key: 4221618 · Received November 3, 2014

Report

Report Number
2024168-2014-07177
Event Type
Malfunction
Date Received
November 3, 2014
Date of Event
October 15, 2014
Report Date
October 15, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P070015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED BALLOON RUPTURE/TEAR WAS CONFIRMED. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. A QUERY OF THE ELECTRONIC COMPLAINT HANDLING DATABASE REVEALED NO OTHER INCIDENTS FOR BALLOON RUPTURE/TEAR REPORTED FROM THIS LOT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION. THOUGH THE DEVICE IS NOT APPROVED FOR SALE IN THE U.S., IT USES A DELIVERY SYSTEM WHICH IS SIMILAR TO A DEVICE SOLD IN THE U.S. THE PRODUCT CODE AND THE PMA# ARE BASED ON THE PREDICATE DEVICE (XIENCE V STENT SYSTEM) THAT IS DETERMINED TO BE SAME AND SIMILAR TO THE DELIVERY SYSTEM OF THIS DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE MID RIGHT CORONARY ARTERY (RCA) WITH MILD CALCIFICATION. PRE-DILATATION WAS PERFORMED, REDUCING THE STENOSIS TO LESS THAN 40%. THE 2.5 X 18 MM IMPLANT DELIVERY SYSTEM WAS ADVANCED TO THE LESION WITHOUT RESISTANCE, BUT WHEN ATTEMPTING TO DEPLOY THE IMPLANT, THE BALLOON WOULD NOT INFLATE AND THE BALLOON WAS OBSERVED TO BE RUPTURED. THREE OTHER IMPLANTS WERE DEPLOYED TO SUCCESSFULLY COMPLETE THE PROCEDURE WITH A GOOD PATIENT OUTCOME. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
703052 ABSORB BIORESORBABLE VASCULAR SCAFFOLD SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 4030461

Patients

Seq Age Sex Outcome Treatment
1