FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
MDR report key: 4221613
·
Received November 3, 2014
Report
- Report Number
- 3015876-2014-01302
- Event Type
- Malfunction
- Date Received
- November 3, 2014
- Date of Event
- October 6, 2014
- Report Date
- October 6, 2014
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K103567
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). PHYSIO-CONTROL EVALUATED THE CUSTOMER¿S DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO-CONTROL REPLACED THE POWER PCB ASSEMBLY. THEREAFTER PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING AND THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.
Description of Event or Problem · 1
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON DURING A PATIENT EVENT. ANOTHER DEVICE THAT WAS AVAILABLE AT THE SCENE OF THE INCIDENT WAS USED TO PROVIDE TREATMENT. THIS ISSUE IS PATIENT RELATED; HOWEVER, THERE WAS NO ADVERSE PATIENT OUTCOME. NO FURTHER DETAILS ABOUT THE PATIENT EVENT OR THE PATIENT WERE PROVIDED BY THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 702257 | LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR | DEFIBRILLATORS, AUTOMATIC, EXTERNAL | MKJ | PHYSIO-CONTROL, INC | 15 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |