FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR

MDR report key: 4221613 · Received November 3, 2014

Report

Report Number
3015876-2014-01302
Event Type
Malfunction
Date Received
November 3, 2014
Date of Event
October 6, 2014
Report Date
October 6, 2014
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K103567
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE CUSTOMER¿S DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO-CONTROL REPLACED THE POWER PCB ASSEMBLY. THEREAFTER PROPER DEVICE OPERATION WAS OBSERVED THROUGH FUNCTIONAL AND PERFORMANCE TESTING AND THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THEIR DEVICE WOULD NOT POWER ON DURING A PATIENT EVENT. ANOTHER DEVICE THAT WAS AVAILABLE AT THE SCENE OF THE INCIDENT WAS USED TO PROVIDE TREATMENT. THIS ISSUE IS PATIENT RELATED; HOWEVER, THERE WAS NO ADVERSE PATIENT OUTCOME. NO FURTHER DETAILS ABOUT THE PATIENT EVENT OR THE PATIENT WERE PROVIDED BY THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
702257 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR DEFIBRILLATORS, AUTOMATIC, EXTERNAL MKJ PHYSIO-CONTROL, INC 15

Patients

Seq Age Sex Outcome Treatment
1