FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4221604 · Received November 3, 2014

Report

Report Number
2032227-2014-46569
Event Type
Malfunction
Date Received
November 3, 2014
Date of Event
October 4, 2014
Report Date
October 4, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. HOWEVER, NO BUTTON ERROR ALARM WAS OBSERVED DURING TESTING. A CRACKED RESERVOIR TUBE LIP, SCRATCHED LCD WINDOW AND SCRATCHED RESERVOIR TUBE WINDOW NOTED.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING A BUTTON ERROR ALARM ON THE INSULIN PUMP THAT THEY WERE UNABLE TO CLEAR. CUSTOMER STATED THAT THE INSULIN PUMP HAS BEEN BUMPED. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 136 MG/DL. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
702246 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523RNAS

Patients

Seq Age Sex Outcome Treatment
1 51 YR