PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-46566
- Event Type
- Malfunction
- Date Received
- November 3, 2014
- Date of Event
- October 4, 2014
- Report Date
- October 4, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- OTHER
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP HAD NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS AND A CRACKED RESERVOIR TUBE LIP.
CUSTOMER REPORTED RECEIVING A BUTTON ERROR ALARM ON THE INSULIN PUMP AND STATED THAT THE KEYPAD WAS UNRESPONSIVE. CUSTOMER'S BLOOD GLUCOSE READINGS AT THE TIME OF THE CALL WERE 53 MG/DL DUE TO NOT EATING. CUSTOMER HAS TREATED WITH A MEAL. NO FURTHER INFORMATION REPORTED.