530G INSULIN PUMP
Report
- Report Number
- 2032227-2014-46696
- Event Type
- Malfunction
- Date Received
- November 3, 2014
- Date of Event
- October 4, 2014
- Report Date
- October 4, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN
- Reporter Occupation
- PATIENT
Narratives
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
(B)(4).CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER REPORTED THE INSULIN PUMP ALARMED MOTOR ERROR DURING BOLUS. CUSTOMER'S BLOOD GLUCOSE WAS 212 MG/DL. CUSTOMER DOES NOT RECALL ANY PREVIOUS SIGNIFICANT EVENT THAT MAY HAVE CAUSED THE DEVICE TO ALARM. THE DEVICE WAS NOT EXPOSED TO MRI. ALARM WAS CLEARED. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE DEVICE. CUSTOMER DOES USE THE SENSOR FEATURE. CUSTOMER WAS ADVISED ABOUT FALSE MOTOR ERROR ALARMS OCCURRING, IF CUSTOMER TRIED TO SEE THE SENSOR GLUCOSE GRAPH WHILE BOLUSING. CUSTOMER WAS ABLE TO REWIND THE DEVICE. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 701685 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-751NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR |