FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 4221428 · Received November 3, 2014

Report

Report Number
1416980-2014-38530
Event Type
Malfunction
Date Received
November 3, 2014
Date of Event
October 9, 2014
Report Date
October 9, 2014
Manufacturer
SHARP CORPORATION OSAKA JAPAN
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. THE DEVICE WAS VISUALLY INSPECTED, FUNCTIONAL AND BATTERY TESTS WERE PERFORMED. DURING SERVICE AN F-94 ALARM WAS IDENTIFIED IN THE ALARM LOG; THE CAUSE WAS DETERMINED TO BE DUE TO LOSS OF CONFIGURATION. TO CORRECT THE CONDITION, THE CONFIGURATION WAS RESET. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING PRODUCT SERVICE BY A SERVICE TECHNICIAN, A FLOGARD INFUSION PUMP HAD AN F-94 ALARM. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
701622 FLOGARD PUMP, INFUSION FRN SHARP CORPORATION OSAKA JAPAN

Patients

Seq Age Sex Outcome Treatment
1