G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2014-25760
- Event Type
- Malfunction
- Date Received
- November 3, 2014
- Date of Event
- October 6, 2014
- Report Date
- October 6, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
(B)(4). THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION. THE COMPLAINT OF NO VIBRATION COULD NOT BE CONFIRMED.
(B)(4).
PATIENT'S FATHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO CLAIM NO VIBRATION PATIENT EXPERIENCED ON (B)(6) 2014. AT THE TIME OF THE CALL TO DEXCOM TECHNICAL SUPPORT, NO MEDICAL INTERVENTION OR INJURIES WERE REPORTED. ADDITIONALLY, DURING THE CALL, DEXCOM TECHNICAL SUPPORT ADVISED PATIENT'S FATHER TO TEST THE ALERT FUNCTIONALITY AND REPORTED DEVICE DID NOT VIBRATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 701859 | G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | MT22430-PNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |