FDA Adverse Event Malfunction Summary report: N

ENSEAL G2 ARTICULATING

MDR report key: 4221325 · Received November 3, 2014

Report

Report Number
3005075853-2014-07557
Event Type
Malfunction
Date Received
November 3, 2014
Date of Event
October 10, 2014
Report Date
October 23, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GEI
PMA / PMN Number
K122797
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. DEVICE EVALUATION - NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RECEIVED IN GOOD PHYSICAL CONDITION. THE DEVICE WAS TESTED ON THE GENERATOR AND PASSED ALL FUNCTIONAL TESTING. THE ENERGY OUTPUT DELIVERED FROM THE DEVICE WAS VERIFIED. ALL THREE TONES WERE HEARD DURING FUNCTIONAL TESTING (TONE 1 IS HEARD WHEN THE ENERGY ACTIVATION BUTTON IS PRESSED; TONE 2 IS HEARD WHEN TISSUE IMPEDANCE THRESHOLD IS REACHED AND TONE 3 IS HEARD WHEN THE CYCLE IS COMPLETE). NO ISSUES WERE NOTED WITH OPENING AND CLOSING OF THE JAWS. THE DEVICE WAS TESTED WITH THE TEST MEDIA AND NO ABNORMAL FORCE WAS NEEDED TO OPEN THE JAWS AFTER THE CUT TEST. THERE WERE NO ANOMALIES NOTED WITH THE FUNCTIONALITY OF THE DEVICE. THIS REPORT IS NOT INTENDED TO DENY THAT YOU EXPERIENCED A PROBLEM WITH THE DEVICE. THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT WE WERE UNABLE TO DUPLICATE DURING OUR LABORATORY ANALYSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A GYNECOLOGICAL PROCEDURE, THE INSTRUMENT LOCKED ON TISSUE AND SIGNIFICANT FORCE WAS NEEDED TO OPEN THE JAWS TO REMOVE THEM FROM THE TISSUE. THE CASE WAS COMPLETED WITH ANOTHER WITH NO CONSEQUENCE TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
700778 ENSEAL G2 ARTICULATING . GEI ETHICON ENDO-SURGERY, LLC. NA L4EX13

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR