JAGWIRE?
Report
- Report Number
- 3005099803-2014-03442
- Event Type
- Malfunction
- Date Received
- November 3, 2014
- Date of Event
- October 6, 2014
- Report Date
- October 9, 2014
- Manufacturer
- BOSTON SCIENTIFIC - COSTA RICA (COYOL)
- Product Code
- EZB
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CN
- Reporter Occupation
- PHYSICIAN
Narratives
REPORTED EVENT OF GUIDEWIRE DISTAL TIP DETACHED EXPOSING THE TIP OF THE METAL COREWIRE. THE DEVICE HAS BEEN RECEIVED FOR ANALYSIS. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
A VISUAL EXAMINATION OF THE GUIDEWIRE REVEALED THAT THE DISTAL TIP WAS DETACHED EXPOSING THE TIP OF THE METAL COREWIRE, APPROXIMATELY 0.4CM . PRESENCE OF ADHESIVE REMNANTS WAS FOUND INDICATING THAT THE PEBAX WAS PROPERLY ATTACHED TO THE COREWIRE. NO EVIDENCE OF COREWIRE FRACTURED. THE COMPLAINT IS CONSISTENT WITH THE RETURN THAT THE DISTAL TIP WAS DETACHED EXPOSING THE TIP OF THE METAL COREWIRE. BASED ON ALL GATHERED INFORMATION, THE MOST PROBABLE ROOT CAUSE IS "HANDLING DAMAGE." SINCE THE MANUFACTURING OF THIS COMPLAINT DEVICE, A CORRECTIVE ACTION HAS BEEN INITIATED TO ADDRESS THIS ISSUE. A DHR (DEVICE HISTORY RECORD) REVIEW WAS PERFORMED AND NO DEVIATION WAS FOUND.
NOTE: THIS REPORT PERTAINS TO ONE OF FOUR DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT # 3005099803-2014-03442, 3005099803-2014-03443, 3005099803-2014-03444 AND 3005099803-2014-03445 FOR THE OTHER ASSOCIATED DEVICE INFORMATION. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE GUIDEWIRE WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHILE THE PHYSICIAN WAS ADVANCING THE JAGWIRE GUIDEWIRE INTO THE PATIENT, THE HYDROPHILIC TIP DETACHED IN THE PAPILLA EXPOSING THE TIP OF THE METAL COREWIRE. THE PHYSICIAN SPENT NEARLY TWO HOURS AND SUCCESSFULLY RETRIEVED THE DETACHED TIP. THREE MORE JAGWIRE GUIDEWIRES WERE OPENED AND DURING PREPARATION, THE HYDROPHILIC TIP DETACHED EXPOSING THE TIP OF THE METAL COREWIRE. THE PROCEDURE WAS COMPLETED WITH DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
NOTE: THIS REPORT PERTAINS TO ONE OF FOUR DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT # 3005099803-2014-03442, 3005099803-2014-03443, 3005099803-2014-03444 AND 3005099803-2014-03445 FOR THE OTHER ASSOCIATED DEVICE INFORMATION. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE GUIDEWIRE WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHILE THE PHYSICIAN WAS ADVANCING THE JAGWIRE GUIDEWIRE INTO THE PATIENT, THE HYDROPHILIC TIP DETACHED IN THE PAPILLA EXPOSING THE TIP OF THE METAL COREWIRE. THE PHYSICIAN SPENT NEARLY TWO HOURS AND SUCCESSFULLY RETRIEVED THE DETACHED TIP. THREE MORE JAGWIRE GUIDEWIRES WERE OPENED AND DURING PREPARATION, THE HYDROPHILIC TIP DETACHED EXPOSING THE TIP OF THE METAL COREWIRE. THE PROCEDURE WAS COMPLETED WITH DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 701181 | JAGWIRE? | STYLET FOR CATHETER, GASTRO-UROLOGY | EZB | BOSTON SCIENTIFIC - COSTA RICA (COYOL) | M00556581 | 16293730 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |