DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
MDR report key: 4221052
·
Received November 3, 2014
Report
- Report Number
- 2938836-2014-17446
- Event Type
- Injury
- Date Received
- November 3, 2014
- Date of Event
- September 3, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE LEAD WAS OTHERWISE NORMAL.
Additional Manufacturer Narrative · 1
DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING ROUTINE FOLLOW-UP FOR THE ASYMPTOMATIC PATIENT, LOSS OF CAPTURE WAS NOTED. X-RAY SHOWED THE RIGHT VENTRICULAR LEAD DISLODGED INTO THE CORONARY SINUS. AN ATTEMPT TO REPOSITION THE LEAD WAS UNSUCCESSFUL. THE LEAD WAS EXPLANTED AND REPLACED WITHOUT COMPLICATION. PATIENT WILL BE MONITORED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 702285 | DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7120Q/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Required Intervention |