FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 4221052 · Received November 3, 2014

Report

Report Number
2938836-2014-17446
Event Type
Injury
Date Received
November 3, 2014
Date of Event
September 3, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE LEAD WAS OTHERWISE NORMAL.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING ROUTINE FOLLOW-UP FOR THE ASYMPTOMATIC PATIENT, LOSS OF CAPTURE WAS NOTED. X-RAY SHOWED THE RIGHT VENTRICULAR LEAD DISLODGED INTO THE CORONARY SINUS. AN ATTEMPT TO REPOSITION THE LEAD WAS UNSUCCESSFUL. THE LEAD WAS EXPLANTED AND REPLACED WITHOUT COMPLICATION. PATIENT WILL BE MONITORED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
702285 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7120Q/65 NA

Patients

Seq Age Sex Outcome Treatment
1 78 YR Required Intervention