FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 4221027 · Received November 3, 2014

Report

Report Number
2938836-2014-17471
Event Type
Injury
Date Received
November 3, 2014
Date of Event
September 17, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION INCLUDED. A PARTIAL LEAD WITH THE CONNECTOR PIN MEASURING 11.2CM WAS RETURNED FOR ANALYSIS. THE PORTION OF THE LEAD THAT WAS RETURNED WAS OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE CLINIC AFTER RECEIVING A VIBRATORY ALERT. HIGH, OUT OF RANGE, PACING LEAD IMPEDANCE AND HIGH CAPTURE THRESHOLD WERE OBSERVED. CLAVICULAR CRUSH WAS NOTED. THE LEAD WAS CAPPED AND REPLACED. PATIENT WAS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
703070 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7121/65 NA

Patients

Seq Age Sex Outcome Treatment
1 50 YR Required Intervention