FDA Adverse Event
Injury
Summary report: N
UNIFY CRT-D
MDR report key: 4221021
·
Received November 3, 2014
Report
- Report Number
- 2938836-2014-17462
- Event Type
- Injury
- Date Received
- November 3, 2014
- Date of Event
- August 26, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NIK
- PMA / PMN Number
- P030054
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.
Additional Manufacturer Narrative · 1
THE REPORTED PREMATURE BATTERY DEPLETION WAS CONFIRMED IN THE LABORATORY. BASED ON ALL AVAILABLE PARAMETER AND USAGE INFORMATION, DEVICE LONGEVITY WAS FOUND TO BE BELOW THE EXPECTED LIMITS. WITH AN EXTERNAL POWER SUPPLY ATTACHED, THE DEVICE FUNCTIONED NORMALLY. NO HIGH CURRENT WAS DETECTED DURING TESTING AND THE POWER CONSUMPTION WAS NORMAL. THE ORIGINAL BATTERY WAS RETURNED TO THE VENDOR FOR FURTHER EVALUATION AND ANALYSIS COULD NOT CONCLUSIVELY DETERMINE A ROOT CAUSE FOR THE PREMATURE BATTERY DEPLETION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WAS EXPLANTED AND REPLACED DUE TO PREMATURE ERI.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 703018 | UNIFY CRT-D | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | NIK | ST. JUDE MEDICAL, INC., CRMD | CD3231-40 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Required Intervention |