FDA Adverse Event Injury Summary report: N

UNIFY CRT-D

MDR report key: 4221021 · Received November 3, 2014

Report

Report Number
2938836-2014-17462
Event Type
Injury
Date Received
November 3, 2014
Date of Event
August 26, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
P030054
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

THE REPORTED PREMATURE BATTERY DEPLETION WAS CONFIRMED IN THE LABORATORY. BASED ON ALL AVAILABLE PARAMETER AND USAGE INFORMATION, DEVICE LONGEVITY WAS FOUND TO BE BELOW THE EXPECTED LIMITS. WITH AN EXTERNAL POWER SUPPLY ATTACHED, THE DEVICE FUNCTIONED NORMALLY. NO HIGH CURRENT WAS DETECTED DURING TESTING AND THE POWER CONSUMPTION WAS NORMAL. THE ORIGINAL BATTERY WAS RETURNED TO THE VENDOR FOR FURTHER EVALUATION AND ANALYSIS COULD NOT CONCLUSIVELY DETERMINE A ROOT CAUSE FOR THE PREMATURE BATTERY DEPLETION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS EXPLANTED AND REPLACED DUE TO PREMATURE ERI.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
703018 UNIFY CRT-D IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3231-40 NA

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention