FDA Adverse Event Injury Summary report: N

FORTIFY VR, DF-4 CONNECTOR

MDR report key: 4220992 · Received November 3, 2014

Report

Report Number
2938836-2014-17435
Event Type
Injury
Date Received
November 3, 2014
Date of Event
September 3, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
LWS
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

THE REPORTED INABILITY TO COMMUNICATE WAS CONFIRMED IN THE LABORATORY AND WAS DUE TO LOW BATTERY VOLTAGE. BASED ON ALL AVAILABLE PARAMETER AND USAGE INFORMATION, DEVICE LONGEVITY WAS FOUND TO BE BELOW THE EXPECTED LIMITS. WITH AN EXTERNAL POWER SUPPLY ATTACHED, THE DEVICE FUNCTIONED NORMALLY. NO HIGH CURRENT WAS DETECTED DURING TESTING AND THE POWER CONSUMPTION WAS NORMAL. THE ORIGINAL BATTERY WAS RETURNED TO THE VENDOR FOR FURTHER EVALUATION AND ANALYSIS COULD NOT CONCLUSIVELY DETERMINE A ROOT CAUSE FOR THE PREMATURE BATTERY DEPLETION.

Description of Event or Problem · 1

IT WAS REPORTED THAT MULTIPLE ATTEMPTS TO INTERROGATE THE DEVICE WERE UNSUCCESSFUL. THE PATIENT RECEIVED A VIBRATORY ALERT ONE MONTH PREVIOUS BUT DID NOT CONTACT THE CLINIC. THE DEVICE WAS EXPLANTED AND REPLACED. PATIENT CONDITION WAS GOOD AFTER THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
702583 FORTIFY VR, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST. JUDE MEDICAL, INC., CRMD CD1233-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention