INSIGNIA
Report
- Report Number
- 2124215-2014-16872
- Event Type
- Injury
- Date Received
- November 3, 2014
- Date of Event
- September 8, 2014
- Report Date
- September 24, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, DETAILED MECHANICAL AND ELECTRICAL TESTING WAS PERFORMED ON THE DEVICE. THE DEVICE BATTERY STATUS WAS END OF LIFE (EOL). THE DEVICE FUNCTIONED NORMALLY THROUGHOUT TESTING. LABORATORY ANALYSIS DETERMINED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION; HOWEVER, THE ESTIMATED LONGEVITY REMAINING VALUE APPEARED TO DECREASE MORE QUICKLY THAN EXPECTED BETWEEN ROUTINE FOLLOW-UPS. FACTORS INFLUENCING THE ESTIMATED LONGEVITY REMAINING CALCULATION INCLUDE PACING RATE, AMPLITUDE, PULSE-WIDTH AND LEAD IMPEDANCE. ANY EVEN SLIGHT CHANGES IN THESE FACTORS WILL IMPACT THE BATTERY CONSUMPTION CALCULATION AND THEREFORE THE REMAINING LONGEVITY ESTIMATE. PLEASE NOTE THAT, DESPITE THE DROP IN ESTIMATED LONGEVITY REMAINING, THE ACTUAL BATTERY CONDITION DID NOT CHANGE SIGNIFICANTLY BETWEEN FOLLOW-UPS. IN SUMMARY, IT WAS DETERMINED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION, BUT DECLARED EOL EARLIER THAN PREVIOUSLY ESTIMATED.
(B)(4). THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS SUSPECTED TO HAVE PREMATURE BATTERY DEPLETION (PBD). UPON CONSULTATION OF THE PATIENT, IT WAS SEEN THAT ON THE PREVIOUS FOLLOW UP, BATTERY REMAINING WAS AT 1.5 YEARS. HOWEVER, BEFORE THE INDICATED BATTERY LIFE ELAPSED, THE DEVICE REACHED END OF LIFE (EOL). ADDITIONAL INFORMATION INDICATED THAT THE DEVICE WAS SUCCESSFULLY EXPLANTED AND REPLACED. THIS DEVICE WAS NO LONGER IN SERVICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
--
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 702672 | INSIGNIA | IMPLANTABLE PULSE GENERATOR | LWP | GUIDANT CRM CLONMEL IRELAND | 1290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| L| R | J174| 1290 |