FDA Adverse Event Injury Summary report: N

QUADRA ASSURA MP CRT-D, DF-4 CONNECTOR

MDR report key: 4220937 · Received November 3, 2014

Report

Report Number
2938836-2014-17432
Event Type
Injury
Date Received
November 3, 2014
Date of Event
September 19, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THE REPORTED FIELD EVENT OF AN INABILITY TO INTERROGATE THE DEVICE WAS NOT CONFIRMED IN THE LABORATORY. THE REPORTED FIELD EVENT OF BACKUP VVI WAS CONFIRMED IN THE LABORATORY VIA REVIEW OF THE DEVICE IMAGE. THE DEVICE WAS TESTED ON THE BENCH AS WELL AS USING AUTOMATED TEST EQUIPMENT AND NO ANOMALIES WERE FOUND. THE ROOT CAUSE OF THE RESET COULD NOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS FOUND TO BE IN BACKUP VVI MODE AND COULD NOT BE INTERROGATED DESPITE SEVERAL ATTEMPTS. THE DEVICE WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
703467 QUADRA ASSURA MP CRT-D, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3371-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention