ETS LINEAR CUTTER
Report
- Report Number
- 3005075853-2014-07565
- Event Type
- Malfunction
- Date Received
- November 3, 2014
- Date of Event
- September 29, 2014
- Report Date
- October 15, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS RESULTS FOUND THAT THE ATS45 DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A TR45W CARTRIDGE NOT LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED FULLY FIRED. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. EVENT COULD NOT BE CONFIRMED AS THE DEVICE PERFORMED WITHOUT ANY DIFFICULTIES NOTED DURING THE FUNCTIONAL TESTING. THE DEVICE HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WAS MET PRIOR TO THE RELEASE OF THE DEVICE.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC APPENDECTOMY PROCEDURE, THE DEVICE DID NOT COMPLETELY FIRE. THIS OCCURRED ON THE FIRST FIRING WITH A WHITE CARTRIDGE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 701060 | ETS LINEAR CUTTER | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | L4F14J |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | RELOAD - TR45W |