FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 4220728 · Received November 3, 2014

Report

Report Number
1416980-2014-38508
Event Type
Malfunction
Date Received
November 3, 2014
Date of Event
October 7, 2014
Report Date
October 9, 2014
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED.  AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR 2367 (AIR IN LINE/SET) OCCURRED DURING USE ON THE HOMECHOICE. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
701239 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE