FDA Adverse Event Injury Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 4220711 · Received November 3, 2014

Report

Report Number
2531779-2014-31228
Event Type
Injury
Date Received
November 3, 2014
Report Date
October 31, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE RETAINED CARTRIDGE WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2014 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE RETAINED CARTRIDGE PASSED VISUAL INSPECTION WITH NO DAMAGE OR DEFECTS NOTED. A FILL TEST WAS COMPLETED WITH NO AIR BUBBLES BEING FORMED INSIDE THE CARTRIDGE. A LEAK TEST WAS PERFORMED WITH NO FAILURES BEING OBSERVED; NO LEAKS WERE OBSERVED FROM THE LUER CONNECTION, O-RINGS, OR ANYWHERE ELSE IN THE CARTRIDGE. THERE WAS NO DEFECT FOUND.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE PATIENT CONTACTED ANIMAS ON (B)(6) 2014 ALLEGING A CARTRIDGE LEAK. THE PATIENT STATED THAT THEY NOTICED THE LEAK PRIOR TO USE AND THERE WAS VISIBLE DAMAGE TO THE CARTRIDGE LUER LOCK. THE PATIENT STATED THAT THEY HAD A BLOOD GLUCOSE (BG) OF 28-29MMOL/L WITH POLYDIPSIA AND TRACE KETONES. THE PATIENT DID NOT RECEIVE ANY TREATMENT ABOVE AND BEYOND THE USUAL ROUTINE CARE OF DIABETES MANAGEMENT. THIS REPORT IS BEING MADE DUE TO THE BG EXCURSION THE PATIENT EXPERIENCED DUE TO AN ALLEGED CARTRIDGE LEAK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
700930 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION C200008

Patients

Seq Age Sex Outcome Treatment
1 21 YR Life Threatening