FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 4220688
·
Received November 3, 2014
Report
- Report Number
- 1416980-2014-38510
- Event Type
- Malfunction
- Date Received
- November 3, 2014
- Date of Event
- October 9, 2014
- Report Date
- October 9, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN AS PART OF PREVENTATIVE MAINTENANCE. THIS IS AN ANCILLARY SERVICE EVENT. THE DEVICE WAS VISUALLY INSPECTED, AND FUNCTIONALLY AND BATTERY TESTED. INSPECTION REVEALED THAT THE FUSE WAS BLOWN; THE CAUSE WAS DUE TO ELECTRICAL DISCHARGE. TO CORRECT THE CONDITION, THE FUSE WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
DURING PRODUCT SERVICE BY A SERVICE TECHNICIAN, A FLOGARD INFUSION PUMP WAS FOUND TO HAVE A BLOWN FUSE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 701337 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |