FDA Adverse Event Malfunction Summary report: N

STELLARIS 25GA POSTERIOR PACK W AFI AND FF 6/BX

MDR report key: 4219248 · Received October 28, 2014

Report

Report Number
1920664-2014-00225
Event Type
Malfunction
Date Received
October 28, 2014
Date of Event
September 30, 2014
Report Date
September 30, 2014
Manufacturer
BAUSCH & LOMB, INC.
Product Code
HQC
PMA / PMN Number
K101325
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVAL COMPLETED RECEIVED ONE BL5425W PACK FROM LOT V2979. THIS PACK CONTAINS A 25GA VITRECTOMY CUTTER. THE TIP PROTECTOR WAS NOT RETURNED. THE CUTTER HAS BEEN TAPED INTO THE PRIMING WELL TRAY AS A MAKE SHIFT TIP PROTECTOR. THE NEEDLE WAS BENT. THE PORT WINDOW WAS IN THE OPENED POSITION. THERE WERE DRIED SOLUTION VISIBLE IN THE TUBING. THE BACK CAP WAS ALIGNED CORRECTLY WITH THE VENT HOLE IN THE SIDE OF THE CUTTER BODY. THE CONNECTORS WERE INSPECTED AND NO DEFECTS WERE FOUND. A FUNCTIONAL TEST WAS PERFORMED USING A STELLARIS PC SYSTEM. THE CUTTER HAD GOOD CLEAN CUTS WITH GOOD ASPIRATION. THE STERILIZATION AND LOT HISTORY RECORDS WERE REVIEWED AND FOUND TO BE ACCEPTABLE. REPORT 3 OF 3. SEE 1920644-2014-00219 AND 00224.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THE VITRECTOMY CUTTER WAS NOT CUTTING AT 4000 CPM, AND WOULD NOT ASPIRATE. THERE WAS NO MEDICAL INTERVENTION REQUIRED AND NO PT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
690455 STELLARIS 25GA POSTERIOR PACK W AFI AND FF 6/BX 25GA POSTERIOR PACK HQC BAUSCH & LOMB, INC. V2979

Patients

Seq Age Sex Outcome Treatment
1