STELLARIS 25GA POSTERIOR PACK W AFI AND FF 6/BX
Report
- Report Number
- 1920664-2014-00225
- Event Type
- Malfunction
- Date Received
- October 28, 2014
- Date of Event
- September 30, 2014
- Report Date
- September 30, 2014
- Manufacturer
- BAUSCH & LOMB, INC.
- Product Code
- HQC
- PMA / PMN Number
- K101325
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
EVAL COMPLETED RECEIVED ONE BL5425W PACK FROM LOT V2979. THIS PACK CONTAINS A 25GA VITRECTOMY CUTTER. THE TIP PROTECTOR WAS NOT RETURNED. THE CUTTER HAS BEEN TAPED INTO THE PRIMING WELL TRAY AS A MAKE SHIFT TIP PROTECTOR. THE NEEDLE WAS BENT. THE PORT WINDOW WAS IN THE OPENED POSITION. THERE WERE DRIED SOLUTION VISIBLE IN THE TUBING. THE BACK CAP WAS ALIGNED CORRECTLY WITH THE VENT HOLE IN THE SIDE OF THE CUTTER BODY. THE CONNECTORS WERE INSPECTED AND NO DEFECTS WERE FOUND. A FUNCTIONAL TEST WAS PERFORMED USING A STELLARIS PC SYSTEM. THE CUTTER HAD GOOD CLEAN CUTS WITH GOOD ASPIRATION. THE STERILIZATION AND LOT HISTORY RECORDS WERE REVIEWED AND FOUND TO BE ACCEPTABLE. REPORT 3 OF 3. SEE 1920644-2014-00219 AND 00224.
THE USER FACILITY REPORTED THE VITRECTOMY CUTTER WAS NOT CUTTING AT 4000 CPM, AND WOULD NOT ASPIRATE. THERE WAS NO MEDICAL INTERVENTION REQUIRED AND NO PT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 690455 | STELLARIS 25GA POSTERIOR PACK W AFI AND FF 6/BX | 25GA POSTERIOR PACK | HQC | BAUSCH & LOMB, INC. | V2979 |