BRAVO
Report
- Report Number
- 9710107-2014-00206
- Event Type
- Malfunction
- Date Received
- September 30, 2014
- Report Date
- September 28, 2014
- Manufacturer
- IVEN IMAGING LTD.
- Product Code
- FFT
- PMA / PMN Number
- K002028
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
(B)(4). EVALUATION SUMMARY: ONE DELIVERY SYSTEM AND CAPSULE WERE RETURNED TO MEDTRONIC FOR EVALUATION. THE DELIVERY SYSTEM WAS INVESTIGATED VISUALLY FOR EXTERNAL DAMAGE. THE TROCAR NEEDLE WAS ADVANCED. THERE WERE NO SIGNS OF BLOOD OR TISSUE ON THE DEVICE. THE DELIVERY SYSTEM WAS NOT BENT AND THE PLUNGER WAS NOT BROKEN. THE EMERGENCY RELEASE ON THE DELIVERY SYSTEM WAS NOT IMPLEMENTED. THE CAPSULE ELECTRODES WERE VISUALLY ACCEPTABLE. THE WIRE THAT HOLDS THE CAPSULE WAS COMPLETELY OFF AND THE FOAM GASKET WAS IN GOOD CONDITION. THE DELIVERY SYSTEM DID NOT HAVE ANY VISIBLE DAMAGE. AS THE PRODUCT WAS RECEIVED, THE DEVICE FUNCTIONED PER SPECIFICATION. THERE WAS NO REQUIRED INTERVENTION TO PREVENT PERMANENT IMPAIRMENT/DAMAGE IN THIS CASE. THE INFORMATION WAS UPDATED IN THIS SUPPLEMENTAL.
A REPEAT PROCEDURE WAS NECESSARY DUE TO THE ALLEGED DEVICE MALFUNCTION. INTERVENTION WAS NOT REQUIRED. THERE WAS NOTHING UNUSUAL ABOUT THE PATIENT OR PROCEDURE ITSELF THAT MAY HAVE LED TO THIS EVENT. AN ENDOSCOPY WAS NOT PERFORMED PRIOR TO THE PROCEDURE. THE DEVICE OPERATOR HAS BEEN PERFORMING THIS PROCEDURE FOR TEN YEAR. NO OTHER KNOWN ADVERSE EVENTS WERE REPORTED.
CUSTOMER REPORTED ON A BRAVO PH CAPSULE THAT FAILED TO ATTACH. THE PT WAS NOT INJURED FOLLOWING THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 606826 | BRAVO | BRAVO PH CAPSULE | FFT | IVEN IMAGING LTD. | FGS-0313 | 24892Q |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |