INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Report
- Report Number
- 9611451-2014-00837
- Event Type
- Malfunction
- Date Received
- October 30, 2014
- Report Date
- October 3, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K103767
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT DEVICE IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE IN NEW ZEALAND FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
(B)(4). METHOD: ONLY ONE RT265 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. IT WAS VISUALLY INSPECTED AND PRESSURE TESTED FOR A LEAK. RESULTS: VISUAL INSPECTION REVEALED NO DAMAGE TO THE SWIVEL ELBOW AND SWIVEL WYE PIECE. THE INSPIRATORY AND EXPIRATORY CONNECTIONS TO THE SWIVEL WYE WERE FOUND TO BE TIGHT. PRESSURE TEST REVEALED THAT THE RETURNED CIRCUIT WAS WITHIN SPECIFICATION. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 140611. CONCLUSION: NO FAULT WAS FOUND WITH THE RETURNED RT265 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT. WE WERE UNABLE TO DETERMINE WHAT MAY HAVE CAUSED THE REPORTED PROBLEM. ALL RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE AND FLOW TESTED DURING PRODUCTION, AND THOSE THAT FAIL ARE REJECTED. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT265 STATE THE FOLLOWING: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE."; "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT."; "SET APPROPRIATE VENTILATOR ALARMS."
A HEALTHCARE FACILITY IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) REPRESENTATIVE THAT THE SWIVEL WYE DISCONNECTED ON TWO RT265 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUITS. ONE WAS FOUND DURING USE AND THE OTHER WAS AFTER PATIENT USE. O PATIENT CONSEQUENCE WAS REPORTED.
A HEALTHCARE FACILITY IN (B)(4) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) REPRESENTATIVE THAT THE SWIVEL WYE DISCONNECTED ON TWO RT265 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUITS. ONE WAS FOUND DURING USE AND THE OTHER WAS AFTER PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.