FDA Adverse Event Malfunction Summary report: N

INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT

MDR report key: 4215910 · Received October 30, 2014

Report

Report Number
9611451-2014-00837
Event Type
Malfunction
Date Received
October 30, 2014
Report Date
October 3, 2014
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
K103767
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT DEVICE IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE IN NEW ZEALAND FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). METHOD: ONLY ONE RT265 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. IT WAS VISUALLY INSPECTED AND PRESSURE TESTED FOR A LEAK. RESULTS: VISUAL INSPECTION REVEALED NO DAMAGE TO THE SWIVEL ELBOW AND SWIVEL WYE PIECE. THE INSPIRATORY AND EXPIRATORY CONNECTIONS TO THE SWIVEL WYE WERE FOUND TO BE TIGHT. PRESSURE TEST REVEALED THAT THE RETURNED CIRCUIT WAS WITHIN SPECIFICATION. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 140611. CONCLUSION: NO FAULT WAS FOUND WITH THE RETURNED RT265 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT. WE WERE UNABLE TO DETERMINE WHAT MAY HAVE CAUSED THE REPORTED PROBLEM. ALL RT265 INFANT DUAL-HEATED EVAQUA2 BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE AND FLOW TESTED DURING PRODUCTION, AND THOSE THAT FAIL ARE REJECTED. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT265 STATE THE FOLLOWING: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE."; "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT."; "SET APPROPRIATE VENTILATOR ALARMS."

Description of Event or Problem · 1

A HEALTHCARE FACILITY IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) REPRESENTATIVE THAT THE SWIVEL WYE DISCONNECTED ON TWO RT265 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUITS. ONE WAS FOUND DURING USE AND THE OTHER WAS AFTER PATIENT USE. O PATIENT CONSEQUENCE WAS REPORTED.

Description of Event or Problem · 1

A HEALTHCARE FACILITY IN (B)(4) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) REPRESENTATIVE THAT THE SWIVEL WYE DISCONNECTED ON TWO RT265 INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUITS. ONE WAS FOUND DURING USE AND THE OTHER WAS AFTER PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.