FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 4214058 · Received October 24, 2014

Report

Report Number
1720753-2014-08897
Event Type
Malfunction
Date Received
October 24, 2014
Date of Event
October 7, 2014
Report Date
October 24, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE X-RAY KEY SWITCH WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED, FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED AN INTERMITTENT KEYSWITCH ERROR MESSAGE ALONG WITH X-RAYS DISABLED MESSAGE. THIS RESULTED IN AN INTERMITTENT, RECOVERABLE LOSS OF IMAGING FUNCTIONALITY. THIS COULD RESULT IN PROCEDURAL DELAYS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
678092 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1