FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 4214035 · Received October 24, 2014

Report

Report Number
1720753-2014-08902
Event Type
Malfunction
Date Received
October 24, 2014
Date of Event
October 7, 2014
Report Date
October 24, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE POWER SUPPLY WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED, FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM TRIPS THE CIRCUIT BREAKER ADJACENT TO THE POWER CORD EVERY TIME IT IS PLUGGED IN. THE FE CONFIRMED THE SYSTEM WOULD NOT BOOT. THIS COULD CAUSE THE DELAY OF A PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
678931 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1