FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 4214035
·
Received October 24, 2014
Report
- Report Number
- 1720753-2014-08902
- Event Type
- Malfunction
- Date Received
- October 24, 2014
- Date of Event
- October 7, 2014
- Report Date
- October 24, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE POWER SUPPLY WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED, FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM TRIPS THE CIRCUIT BREAKER ADJACENT TO THE POWER CORD EVERY TIME IT IS PLUGGED IN. THE FE CONFIRMED THE SYSTEM WOULD NOT BOOT. THIS COULD CAUSE THE DELAY OF A PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 678931 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |