FDA Adverse Event Malfunction Summary report: N

TOTAL CARE BED

MDR report key: 4213913 · Received October 6, 2014

Report

Report Number
1824206-2014-02400
Event Type
Malfunction
Date Received
October 6, 2014
Date of Event
September 8, 2014
Report Date
September 8, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K962942
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECH FOUND THE BED'S SCALE HAD BEEN READING INACCURATE WEIGHTS. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2014. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECH CALIBRATED THE SCALE TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE SCALE WAS READING INACCURATE WEIGHTS. THE BED WAS LOCATED IN THE CCU AT THE ACCOUNT. THERE WAS NO PT/INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).