FDA Adverse Event Malfunction Summary report: N

CATGUT CHROMIC 2/0 (3.5) 100M

MDR report key: 4213905 · Received October 6, 2014

Report

Report Number
2916714-2014-00773
Event Type
Malfunction
Date Received
October 6, 2014
Report Date
October 6, 2014
Manufacturer
B. BRAUN SURGICAL S.A.
Product Code
GAL
PMA / PMN Number
K991223
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

US REPORTING AGENT NOTIFIED ON: (B)(4) 2014. MFG SITE EVAL: SAMPLES RECEIVED: 1 OPEN CASSETTE. THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE/BATCH. THERE ARE (B)(4) UNITS IN OEM STOCK. TESTED THE KNOT PULL TENSILE STRENGTH OF THE CASSETTE RECEIVED AND THE RESULTS FULFILL THE REQUIREMENTS OF THE OEM. REVIEWED THE BATCH MFG RECORD, THIS PRODUCT HAD A NORMAL PROCESS AND THE RESULTS DURING THE PROCESS FULFILL OEM REQUIREMENTS. FINAL CONCLUSION: THE COMPLAINT IS NOT CORRESPONDING (NOT JUSTIFIED). ACTIONS ON PRODUCT: NOT APPLICABLE. CORRECTIVE/PREVENTIVE ACTIONS: NOT APPLICABLE.

Description of Event or Problem · 1

COUNTRY OF COMPLAINT: (B)(6). THREAD TEARS TOO EASILY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
625161 CATGUT CHROMIC 2/0 (3.5) 100M ABSORBABLE SUTURE GAL B. BRAUN SURGICAL S.A. F1154052 613514

Patients

Seq Age Sex Outcome Treatment
1 Other