FDA Adverse Event Malfunction Summary report: N

TRANSTAR STRETCHER

MDR report key: 4213889 · Received August 14, 2014

Report

Report Number
3006697241-2014-00615
Event Type
Malfunction
Date Received
August 14, 2014
Date of Event
July 22, 2014
Report Date
July 22, 2014
Manufacturer
HILLROM DE MEXICO S DE RL DE CV
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE BRAKE AND STEER LINKAGE INOPERABLE. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE HILL-ROM TECHNICIAN THE BRAKE AND STEER LINKAGE TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES ARE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
488119 TRANSTAR STRETCHER WHEELED STRETCHER FPO HILLROM DE MEXICO S DE RL DE CV 8050

Patients

Seq Age Sex Outcome Treatment
1