FDA Adverse Event Malfunction Summary report: N

AFFINITY 4 BED

MDR report key: 4213874 · Received October 7, 2014

Report

Report Number
1824206-2014-02405
Event Type
Malfunction
Date Received
October 7, 2014
Date of Event
September 8, 2014
Report Date
September 8, 2014
Manufacturer
HILL-ROM, INC.
Product Code
HDD
PMA / PMN Number
K915779
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE RELAY ON THE LOGIC BOARD. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE LOGIC BOARD AND THE ISSUE WAS RESOLVED. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE BED HILOW WOULD SELF-RUN WHEN THE BED IS PLUGGED IN. THE BED WAS LOCATED IN OB ROOM 10. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
630003 AFFINITY 4 BED BIRTHING BED HDD HILL-ROM, INC. 3700

Patients

Seq Age Sex Outcome Treatment
1