FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4213527 · Received October 25, 2014

Report

Report Number
2032227-2014-43489
Event Type
Malfunction
Date Received
October 25, 2014
Date of Event
September 26, 2014
Report Date
September 26, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE FUNCTIONAL TESTING. HOWEVER, THE INSULIN PUMP ALARMED FOR RESET ERROR AFTER THE BATTERY WAS CHANGED DUE TO A FAULTY CONNECTOR ON THE INTERFACE BOARD. A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND MINOR SCRATCHES ON THE DISPLAY WINDOW WERE NOTED DURING THE VISUAL INSPECTION.

Description of Event or Problem · 1

CUSTOMER REPORTED THE INSULIN PUMP ALARMED UNEXPECTED RESTART EACH TIME A NEW BATTERY WAS INSERTED. CUSTOMER'S BLOOD GLUCOSE WAS 123 MG/DL. A TEMP BASAL WAS NOT PROGRAMMED BEFORE THE ALARM OCCURRED. THE DEVICE WAS NOT STORED FOR A LONG PERIOD OF TIME. NO FURTHER INFO REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
680641 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP OYC MEDTRONIC MINIMED MMT-523RNAH

Patients

Seq Age Sex Outcome Treatment
1 68 YR