PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-43489
- Event Type
- Malfunction
- Date Received
- October 25, 2014
- Date of Event
- September 26, 2014
- Report Date
- September 26, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP PASSED THE FUNCTIONAL TESTING. HOWEVER, THE INSULIN PUMP ALARMED FOR RESET ERROR AFTER THE BATTERY WAS CHANGED DUE TO A FAULTY CONNECTOR ON THE INTERFACE BOARD. A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND MINOR SCRATCHES ON THE DISPLAY WINDOW WERE NOTED DURING THE VISUAL INSPECTION.
CUSTOMER REPORTED THE INSULIN PUMP ALARMED UNEXPECTED RESTART EACH TIME A NEW BATTERY WAS INSERTED. CUSTOMER'S BLOOD GLUCOSE WAS 123 MG/DL. A TEMP BASAL WAS NOT PROGRAMMED BEFORE THE ALARM OCCURRED. THE DEVICE WAS NOT STORED FOR A LONG PERIOD OF TIME. NO FURTHER INFO REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 680641 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | OYC | MEDTRONIC MINIMED | MMT-523RNAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |